Technical Information
Inlet and exhaust connections
5
MS40+ Single Stage Rotary Vane Pumps User Manual / 87-901-008-01 (H.00) 70/104
NOTE
After switching off the pump using the main switch, wait at least 10 s before
switching it back on.
NOTE
If rigid piping is used, it is good practice to use a flexible joint in order to avoid
undue forcing of the connection on the pump.
The exhaust duct must be connected to a pipe that will take away the
pumped out gases.
NOTE
An internal oil mist eliminator avoids pollution of the surrounding atmosphere by
the oil present in the exhaust duct during pump operation.
CAUTION!
Never block the pump exhaust line. This would cause overpressure in the casing
with the risk of breaking the oil tank.