32 Installation Note E4440-90613
Installation Procedure
17.Refer to Figure 25. Remove the screw from the cable hold down rod on the mid web to
allow the “10” cable and other cables to be free.
Figure 25 Assembly and Cable Location
18.In order to remove the RYTHM end of the cable, and replace the cable, several board
assemblies must be removed. Refer to Figure 25. Please note the position and routing
of all cables. All of the flexible coax cables are color coded, but the ribbon cables are
not.
19.Remove the Front End Driver being careful to avoid damaging the ribbon cables.
20.Remove the Synthesizer assembly after removing the rigid cable and the two
retaining screws.
21.Remove the Reference assembly.