EasyManua.ls Logo

Agilent Technologies U1251A - Adjustment Is Recommended

Agilent Technologies U1251A
168 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Tests and Calibration 7
Agilent U1251A/U1252A User’s and Service Guide 123
warranted beyond the 1- year calibration interval. Agilent
does not recommend extending calibration intervals beyond
2 years for any application.
Adjustment is Recommended
Specifications are only guaranteed within the period stated
from the last adjustment. Whatever calibration interval you
select, Agilent recommends that complete re- adjustment
should always be performed at the calibration interval. This
will assure that the U1251A/U1252A will remain within
specification for the next calibration interval. This criteria
for re- adjustment provides the best long –term stability.
Performance data is measured during Performance Verifica-
tion Tests and this does not guarantee the instrument will
remain within these limits unless the adjustments are per-
formed.
Refer to the “To Read the Calibration Count” on page 153
and verify that all adjustments have been performed.

Table of Contents

Other manuals for Agilent Technologies U1251A

Related product manuals