8 Agilent U1251A/U1252A User’s and Service Guide
Product Regulations
Performance Criteria
1
IEC 61326-1:1997+A1:1998 / EN 61326-1:1997+A1:1998 U1251A U1252A
CISPR 11:1990 / EN 55011:1991 – Group 1 Class A
IEC 61000-4-2:1995+A1:1998 / EN 61000-4-2:1995 (ESD
4kV CD, 8kV AD)
B B
IEC 61000-4-3:1995 / EN 61000-4-3:1995 (3V/m, 80% AM) A A
IEC 61000-4-4:1995 / EN 61000-4-4:1995 (EFT 0.5kV line-
line, 1kV line-earth)
N/A A
IEC 61000-4-5:1995 / EN 61000-4-5:1995 (Surge 0.5kV line-
line, 1kV line-earth)
N/A A
IEC 61000-4-6:1996 / EN 61000-4-6:1996 (3V, 0.15~80 MHz,
80% AM, power line)
N/A A
IEC 61000-4-11:1994 / EN 61000-4-11:1994 (Dips 1 cycle,
100%)
N/A A
Canada: ICES-001:1998
Australia/New Zealand: AS/NZS 2064.1
EMC
IEC 61010-1:2001 / EN 61010-1:2001
Canada: CSA C22.2 No. 61010-1:2004
Safety
USA: UL 61010-1: 2004
Additional Information:
The product herewith complies with the essential requirements of the Low Voltage Directive 73/23/EEC and the
EMC Directive 89/336/EEC (including 93/68/EEC) and carries the CE Marking accordingly (European Union).
1
Performance Criteria:
A Pass - Normal operation, no effect.
B Pass - Temporary degradation, self recoverable.
C Pass - Temporary degradation, operator intervention required.
D Fail - Not recoverable, component damage.
N/A – Not applicable
Notes:
Regulatory Information for Canada
ICES/NMB-001:1998
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est confomre à la norme NMB-001 du Canada.
Regulatory Information for Australia/New Zealand
This ISM device complies with Australian/New Zealand AS/NZS 2064.1