Alcatel Vacuum Technology France --- IPUP A100 Instruction manual
Electrical connection
ELECTRICAL CONNECTION
Electrical cable
RT1 Electrical switch
Female connector
Remove bracket before disconnect-
ing and install it after connection
The pump is supplied with an electrical cable [4 ´ 14 AWG leads
(2.08mm
2
)] (2.5m length) equipped with a female connector.
Connect the green lead to the ground on the connector, the other 3
leads may be connected with the 3 phases.
The pump will automatically correct with autophase correction fixture.
Connect the female connector to the pump male connector and lock it
as follows:
l Screw the ferrule.
l Fit the pipe clamp on the female connector collar (position the fixing
holes correctly).
l Secure the two M3 fixing screws.
2 fixing screws M3
Pipe clamp
Female connector
Ferrule
G
B
0
0
5
2
8