D 70
1 / 1
Alcatel Vacuum Technology France --- IPUP A100 Instruction manual
Pump does not start.
Display is showing a message or
an alarm
Warning
then
Hazard D01: EXHAUST PRESSURE
W01: EXHAUST PRESSURE
NO
Hand held remote control
display is showing
Message on display
STOCK
WAITING ??
INCIDENT CAUSE TROUBLESHOOTING
NO
Exhaust pressure ² 2500 mbar
l Check for clogged
exhaust
Other problem
l Call your Alcatel Service
Representative.
NO
Controller is in slave mode
l Switch RT1 to ”0” for 10s,
then switch it on again.
Other problem
l Call your Alcatel Service
Representative.
NO
Pump in storage mode
l Switch RT1 to ”0”.
Then switch it on again.
Other problem
l Call your Alcatel Service
Representative.
IPUP ADP 100P ONLY