D 100
1 / 1
Alcatel Vacuum Technology France --- IPUP A100 Instruction manual
Fault message display ---
Pressure problem
IPUP A100P WITH HAND HELD
DISPLAY MODULE
Pump is running and displays a
warning
warning
W01: EXHAUST PRESSURE
INCIDENT CAUSE TROUBLESHOOTING
NO
Exhaust pressure > 1700 mbar
l Check the N2 flow rate
setting value. Check that the
exhaust pipe is not clogged.
Other problem
l Call your Alcatel Service
Representative.
Pump is stopped and displays a
hazard
Hazard
D01: EXHAUST PRESSURE
INCIDENT CAUSE TROUBLESHOOTING
NO
Exhaust pressure > 2500 mbar
l Check the N2 flow rate
setting value. Check that the
exhaust pipe is not clogged.
Other problem
l Call your Alcatel Service
Representative.