WARNING
B 50
Alcatel Vacuum Technology France --- IPUP A100 Instruction manual
1 / 2
Electrical connection
l Make sure that breaker RT1 is off during electrical connection.
l Study the preliminary precautions (See B 00).
All the internal electrical connections required for use the pump are
made prior to shipment.
However, the electrical connection of the main power supply is
provided by the process tool.
An out of phase condition is corrected automatically by the pump
electronics.
IPUP A100 does not include any operator replaceable fuses.
PROTECTIVE EARTH
CONNECTION
An IEC 417q5019 symbol is located inside the enclosure,
under the monitoring cooling plate near the main protective
conductor terminal.
PROCESS TOOL ELECTRICAL
PROTECTION
The pump is not equipped with a 10000 AIC main breaker.
Electrical shock may result in severe injury.
Power the pump from a facilities supplied 15 A min main
circuit breaker with a 10000 AIC rating. The user-supplied
circuit breaker should be in close proximity to the pump (no
further than 25 ft within line of sight of the pump).