Alltion (Guangxi) Instrument Co., Ltd.
14
Product Assembly | AC-5000 User’s Manual
Installation Confirmation
Check whether the threaded connection of all assembled positions have been tightened,
judge whether the whole machine has completed installation;
Loosen all locking knobs, check if the rotation and movement of all joints of the whole
machine are smooth, judge whether the mechanical function is normal;
Check whether the locking knob of each joint is valid, judge whether the locking function is
normal;
Turn on the power switch, observe whether the indicator of the power switch turns on,
whether the light is illuminated from the microscope body, judge whether the power is
supplied;
when the dimming knob is adjusted to the minimum, there is no illumination light;
Turn the dimming knob to observe whether the lighting spot changes bright and dark, judge
whether the dimming function is normal;
If the above functions are normal, it can be confirmed that the product is well installed.