Alltion (Guangxi) Instrument Co., Ltd.
Safety Caution | AC-5000 User’s Manual
Requirements on Installation
The installation of the product will be completed by our service representative or by
professional personnel authorized by us. Please make sure that the following operational
requirements have been met:
All the safety-related mechanical connections (please find the Manual for details) are
properly connected and all screws have been tightened.
All wires and plugs work normally.
The adopted power wires meet the design requirements of this instrument.
Please do not place the equipment in a difficultly operation position.
Operating Requirements
Please pay special attention to the PROMPT symbols on the instrument ( especially
Warning sign).
Avoid to watch the light source of the microscope directly, for example, the objective lens of
the microscope.
Do not disassemble and assemble the binocular head barrel and objective lens during use
to prevent falling and injuring the patient.
Do not disassemble or assemble the supporting components during use to prevent the
Support system from being out of balance, damaging the components or hurting the
patient.
This surgical microscope cannot be used for ophthalmic examination and surgery.
Safety Signs on the Instrument
1 Product label
It contains basic information of the product.
2 Warning
Make sure that the binocular tube is safely installed and the
screw has been tightened, otherwise the binocular tube has the
risk of falling.
3 Warning
Make sure the screw has been tightened, Otherwise the cross
arm bracket has the risk of lifting.
4 Load-bearing tip of cross arm bracket
The balance arm would fail in case of an added weight of greater
than 1.5 kg.