24
Operation of the Microscope | AC-5000 User’s Manual
Optical Adjustment of Colposcope
Adjust the colposcope to the minimum magnification, move the colposcope to the
chosen position till the object is observed clearly.
Adjust the pupillary distance of the binocular barrel, and when the images of the two
eyepieces merge into one, the correct position is reached.
Adjust the colposcope to the maximum magnification and move the colposcope to the
position that has clearest image.
Clearer image can be obtained by fine adjustment of the focus knob on the L-shaped.
Change the colposcope to the minimum magnification, and adjust the knob of diopter on
the eyepiece until the clearest image is observed.
User with normal version: adjust the diopter to 0;
User with abnormal version and wearing glasses: adjust the diopter to 0;
User with abnormal version and not wearing glasses: adjust the diopter till gaining
clearest image.
Finish optical adjustment.