Alltion (Guangxi) Instrument Co., Ltd.
AC-5000 User’s Manual | Maintenance related information
Maintenance Related Information
Troubleshooting
Power cable is not connected
Power switch is not turned on
The dimming knob is adjusted to
minimum position
Instrument electrical failure
Contact the local dealer or
after-sales service agent
Contact the local dealer or
after-sales service agent
Illumination fails
intermittently during
use
The cooling window and the air inlet
are covered or blocked by external
object
Remove the foreign object and
clean the cooling window
Contact the local dealer or
after-sales service agent
Instrument electrical failure
Contact the local dealer or
after-sales service agent
Camera switch is not turned on
Instrument electrical failure
Contact the local dealer or
after-sales service agent
The instrument
cannot stop at any
time when it moves
up and down
Balanced arm is not adjusted to
balance after adding or decreasing
accessories of the microscope
Contact the local dealer or
after-sales service agent
The instrument is
running stiffly
The friction adjustment knob is
adjusted too tight.
Loosen the friction adjusting
knob, and moderately adjust
the friction.
Optical
magnification
switching failure
Mechanical failure of the instrument
Contact the local dealer or
after-sales service agent
The built-in camera switch is not
turned on
The input channel of the video cable
is not consistent with the display
channel of the display.
Change the display channel of
the video cable or the display
channel of the display.
Video cable is not connected
correctly
Reconnect the video cable
There is problem for the video cable