2-2 | 3050-SLR Moisture Analyzer
Ranges
0.10 to 100 parts per million by volume (ppmv)
Readings can also be displayed in units of ppmw, lb/mmscf, and mg/Nm
3
.
Outputs
Data:
Isolated 4 to 20 mA, 100 to 500Ω Analog output (software conigurable) RS-232 or RS-485
serial port, two and four wire mode
Alarms/Alerts
Two independent hermetically sealed reed type contact closures 60 VDC, 30 VAC, 10 VA
maximum resistive for system alarm and data valid. All are fail-safe by default.
Alarms are available on RS-485 interface.
Lower Detectable Limit
0.10 PPMV nominal
Sensitivity
0.01 PPMV or 1% of reading, whichever is greater
Accuracy
+/- 0.03 ppmv or +/- 10% of reading, which ever is greater
Reproducibility
+/- 0.015 ppm or +/- 5% of reading, which ever is greater
Moisture Generator Value
1 ppmv nominal

Table of Contents