User Interface | 4-9
To Save settings on the Setup tab, click Apply.
Veriication Tab
Use the Veriication tab to schedule routine Zero.
Verify Now Button
Click to begin veriication cycle now.
Abort Button
Click to terminate veriication or zero.
Zero Now Button
Click to begin zero cycle now.
Adjust Span After Verfy check box
Checked The analyzer performs an span adjustment at the end of
the zero cycle.
Not checked The analyzer veriies only.
Ignore Span Drift check box
Checked One time span value change is not limited.
Not checked One time span value change is limited to 10% of the
existing span value.
Set Clock
Click Set Clock to synchronize the clock within the analyzer with the PC. The Time Syn-
chronization box opens with the PC time and date and the analyzer time and date. Click
Synchronize to set the time, or click Cancel to close the box.
Figure 4-6.
Veriication tab.