EasyManuals Logo

Ametek 3050 Series User Manual

Ametek 3050 Series
86 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #80 background imageLoading...
Page #80 background image
A-14 | 3050-SLR Moisture Analyzer
All of the coniguration parameters of the analyser can be modiied by the “write one or
multiple holding registers” command. In most cases it is a one step operation involving set-
ting the contents of the corresponding register. Ten different examples of using this com-
mand are presented below. The irst example is presented with all “request” and “response”
formatting information. For brevity, the remaining examples list just the key register infor-
mation.
Example 1: Alarm Enable
TASK.
Enable Alarm Output. Device address is 2.
ACTION.
Write one holding register (function 06). Register address = 102 (holding register #40103).
Value = 1(1-enable, 0-disable).
MODBUS transaction:
Request
Device Address =02Hex
Function Code =06Hex
Register Address Hi =00Hex
Register Address Lo =66Hex
Register Value Hi =00Hex
Register Value Lo =01Hex
CRC Hi
CRC Lo
Response
Device Address =02Hex
Function Code =06Hex
Register Address Hi =00Hex
Register Address Lo =66Hex
RegisterValue Hi =00Hex
Register Value Lo =01Hex
CRC Hi
CRC Lo

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek 3050 Series and is the answer not in the manual?

Ametek 3050 Series Specifications

General IconGeneral
BrandAmetek
Model3050 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals