26 17-08-2004 123312 03
The calibrator will now start working towards the T
max
using the slope
rate selected in SETUP. The flashing SWITCH TEST symbol indicates
the current status:
°
°C
The instrument will check for changes in the SWITCH TEST. If no
change has been detected by the time T
max
is reached, the instrument
will register an ERROR.
The calibrator will stabilise at this temperature, and then work towards
the T
min
using the slope rate selected in SETUP.
The flashing SWITCH TEST symbol indicates the current status:
°
°C
The instrument will check for changes in the SWITCH TEST input once
again. If no change has been detected by the time the T
min
has been
reached, the instrument will register an ERROR.
The results of the test will be displayed as 3 values: an “Open”
temperature, a “Close” temperature and a “Hyste” hysteresis
temperature (the difference between the two temperatures).
The open temperature is shown first:
°C
Press
to display the close temperature:
°C