44 17-08-2004 123312 03
8.2 Adjusting and calibrating the instrument
You are advised to return the calibrator to Ametek Denmark A/S or an
accredited laboratory at least once a year for calibration and
adjustment.
Alternatively, you can calibrate/adjust the calibrator yourself. You will
need a reference thermometer and a reference sensor with a traceable
certificate. Please follow the instructions given below.
Connect the calibrator to an external precision instrument (e.g. a DTI)
as shown in Fig. 11:
Fig. 11
Hold down the
button while pressing the on/off power
control switch.
The instrument is now in adjustment/service mode.
C
A
L
I
B
R
A
T
I
O
N
I
N
S
T
R
U
M
E
N
T
S