123312 03 17-08-2004 47
Press
to accept your selection. The day will be
adjusted if necessary to ensure the legality of the date.
Finally, the day, month and year will flash:
Press
to accept the date.
or
press
to cancel the whole selection.
8.2.2 Calibrating/adjusting the instrument
The internal calibration/adjustment is a complex function which is
divided into a number of different steps:
The instrument will disclose the first calibration temperature by
displaying the text “TEMP.1 XXX°C” for approx. 1 second:
Calibration temperature for calibrators:
ITC-155 A 1. -23°C / -9.4°F
2. 20°C / 68°F
3. 60°C / 140°F
4. 100°C / 212°F
5. 155°C / 311°F