48 17-08-2004 123312 03
ITC-320 A 1. 50°C / 122°F
2. 120°C / 248°F
3. 180°C / 356°F
4. 250°C / 482°F
5. 320°C / 608°F
ITC-650 A 1. 50°C / 122°F
2. 200°C / 392°F
3. 350°C / 662°F
4. 500°C / 932°F
5. 650°C / 1202°F
The instrument will now heat up/cool down to reach the first calibration
temperature:
°C
°
Once the calibrator is stable, you need to enter the reference
temperature found using the reference thermometer. The
calibration temperature is suggested as a reference point:
°C
This procedure is repeated for TEMP.2, TEMP.3, TEMP.4 and
TEMP.5.
All five calibration temperatures and associated reference
temperatures have now been entered.
The instrument will now check whether the reference temperatures
which have been entered are within the permitted tolerances.
Permitted tolerances:
• ITC-155 A : ±0,1°C / 0.18°F
• ITC-320 A : ±0,15°C / 0.27°F
• ITC-650 A : ±0,20°C / 0.36°F