Likely cause: An error has occurred in the control circuit.
Solution: The calibrator should be returned to the
manufacturer for service.
Nothing happens when the power control switch (on/off switch)
is pressed.
Likely cause: There is no power to the calibrator.
Solution: Check that the calibrator is correctly connected.
Check the fuse.
If there are no problems with the mains cable or the
fuse, the calibrator should be returned to the
manufacturer for service.