54 17-08-2004 123312 03
THERMAL SPECIFICATIONS ITC-155 A
80 mm / 3.15 inch axial homogeneity :
0.8°C / 1.44°F @155°C / 311°F
0.5°C / 0.9°F @-20°C / -4°F
Difference between holes :
0.03°C / 0.05°F
Influence from load :
0.15°C / 0.27°F @155°C / 311°F
0.10°C / 0.18°F @-20°C / -4°F
Influence from load with
ext. reference :
0.02°C / 0.03°F @155°C / 311°F
0.02°C / 0.03°F @-20°C / -4°F
Difference between inserts :
0.08°C / 0.14°F
Long term drift (1 year) :
±0.10°C / ±0.18°F
Calibration accuracy (test limit) ±0.10°C / ±0.18°F
Temperature coefficient 0.02°C/°C (0-40°C) / 0.03°F/°F (32-104°F)
Stability ±0.03°C / ±0.05°F
Reference accuracy ±0.10°C / ±0.18°F
Total accuracy ±0.25°C / 0.45°F
Heating time incl. insert -20°C / -4°F to 23°C / 73.4°F : 4 min.
23°C / 73.4°F to 155°C / 311°F : 14 min.
-20°C / -4°F to 155°C / 311°F : 18 min.
Additional stabilisation time 10 min.
Cooling time incl. insert 155°C / 311°F to 100°C / 212°F : 4 min.
155°C / 311°F to 23°C / 73.4°F : 14 min.
23°C / 73.4°F to -20°C / -4°F : 23 min.
155°C / 311°F to -20°C / -4°F : 37 min.