Questions and Answers

  • D
    Danny WeberAug 13, 2025
    What to do if Analyticon Combi Scan 100 Measuring Instruments shows 'Printer out of paper'?
    • B
      bradshawrichardAug 13, 2025
      If the Analyticon Measuring Instruments indicates 'Printer out of paper', refill the printer and try printing again.
  • B
    Brian JonesAug 17, 2025
    What to do if Analyticon Measuring Instruments says 'No strip'?
    • H
      Heather JonesAug 17, 2025
      If the Analyticon Measuring Instruments displays a 'No strip' message, place a strip on the strip holder before pressing the start button!
  • C
    crobertsAug 20, 2025
    What to do if Analyticon Measuring Instruments displays 'Worklist full'?
    • K
      Kyle LeAug 20, 2025
      If the Analyticon Measuring Instruments displays 'Worklist full', it means there are more than 20 data entries in the received worklist. Remove the records above 20 from the transfer list and try downloading again.
  • C
    Christopher SmithAug 25, 2025
    How to fix 'Tray error' on Analyticon Measuring Instruments?
    • S
      Stacey GreenAug 25, 2025
      A 'Tray error' on the Analyticon Measuring Instruments may occur if the tray is missing or its movement is blocked. Check the tray and restart the reader.
  • O
    omargibsonAug 28, 2025
    What to do if Analyticon Combi Scan 100 Measuring Instruments displays 'Calibration error'?
    • G
      Gregory GomezAug 28, 2025
      If the Analyticon Measuring Instruments shows a 'Calibration error', it could be due to a problem with the calibration pad on the strip holder tray. Clean the pad or use a new strip holder tray and restart the reader.
  • D
    Diana TaylorAug 31, 2025
    How to resolve 'Codekey error' on Analyticon Combi Scan 100?
    • C
      colin50Aug 31, 2025
      A 'Codekey error' on the Analyticon Measuring Instruments indicates that the codekey plug is either missing or incorrectly connected. Check if the plug is connected properly, then restart the reader.
  • J
    Jason AdamsSep 5, 2025
    What to do if Analyticon Combi Scan 100 Measuring Instruments displays 'Worklist frame error'?
    • G
      griffithannSep 5, 2025
      If the Analyticon Measuring Instruments displays a 'Worklist frame error', it indicates an issue with the received worklist from the PC. Check the communication cable or the communication settings.
  • D
    Dana EnglishSep 8, 2025
    How to fix 'Netware error' on Analyticon Combi Scan 100?
    • A
      Ariel BrownSep 8, 2025
      A 'Netware error' on the Analyticon Measuring Instruments suggests a problem in bidirectional serial port communication. Check the communication cable or the communication settings.

Analyticon Combi Scan 100 Specifications

General IconGeneral
Data Storage2000 results
DisplayLCD
ThroughputUp to 120 tests per hour
InterfacesRS232, USB
Power Supply100-240V AC, 50/60Hz
Weight2.5 kg
Measurement PrincipleReflectance photometry
Measurement RangeVaries by parameter