6. Warnings / Precautions
The CombiScan® 100 must be used in the manner described and specified in this manual in
order to provide the saftety and performance standards specified.
The instrument is calibrated for
CombiScreen® 11SYS test strips. With other CombiScreen
test strips or with test trips of other manufacturers, wrong results will be obtained.
In order to establish a final diagnosis and prescribe an appropriate therapy, the results obtained
with CombiScan® 100 should be verified with other medical and diagnostic results.
To check the performance of the system (instrument+test strips) and to ensure reliable results,
it is recommended to use control solutions regularly and to record the results. Especially in the
following situations, the use of urine controls is highly recommended:
a) at the beginning of the daily routine measurements
b) if a new bottle of test strips is opened
c) if test results are in doubt
d) if operators are trained on the system
Urine samples should be handled as a potentially infectious liquid. The equipment which comes
into contact with urine, should be disinfected thoroughly after having finished working.
Always apply safe laboratory practices when operating the CombiScan® 100
The CombiScan® 100 is an electrical device. Take care that no liquids like water, urine or disin-
fectants enter the instrument, because this may damage the device or may lead to a dangerous
situation for the user.
Do not open the housing of the CombiScan® 100 when connected to power supply. In case of
problems, call your distributor.
Protect the instrument from extreme temperatures or high atmospheric humidity, and keep it
out of bright light! For measuring, keep the specifications concerning temperature (15 - 30°C)
and relative atmospheric humidity (<85%).