12
Operation
12
Operation
4.4.3 Settings of date and time
4.4.3.1. Adjust date and time
The built-in digital clock enables to store the date and time of the measurement. To get correct
information it is necessary to set the clock exactly. The clock is working even when power sup-
ply is not connected to the equipment.
You can adjust the time in the Menu/Setup/Clock/Set sub-menu. The time or date unit high-
lighted by the cursor can be increased or decreased by pressing the Up or Down buttons. Press-
ing the Right button (>>>) moves the blinking cursor to the right. When the blinking cursor
reached the right corner of the display by pressing the Right button (>>>) again the new
time will be set and Main Menu appears on the display.
4.4.3.2. Changing the display format of date and time
To adjust the display format according to your individual requirements the CombiScan
®
100
supports different date and time formats. You can change the display mode of date and time
after entering the Menu/Setup/Clock/Display sub-menu. Pressing the Left button (Sequence)
allows the date format to be changed to Year/Month/Day or Day/Month/Year or Month/Day/
Year. You can choose between the 24-hour or the 12-hour plus a.m/p.m system by pressing the
“12/24” button.
4.5. Measuring mode
The CombiScan
®
100 can be operated in two different modes:
In Normal mode, the equipment waits for the strip to incubate before it reads the first pad.
Throughput in this mode is 40 tests per hour and in this case incubation is controlled by the
instrument.
In Fast mode, the test strip is measured directly after Start is pressed. In this case it is up to the
user to time the incubation period outside the analyzer. Using Fast mode enables a throughput
of 120 strips per hour.
2004/01/01 12:12:25 000
Worklist ChkMeas Menu
Menu
Setup MeasMode Memory
Measuring Mode: Norm
Normal Fast OK

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Analyticon Combi Scan 100 and is the answer not in the manual?

Analyticon Combi Scan 100 Specifications

General IconGeneral
BrandAnalyticon
ModelCombi Scan 100
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals