OPERATING INSTRUCTIONS
PRE-TEST CONDITIONS
1. The operating temperature range is between -10°C to 50°C. Using the device within this
temperature range will limit condensation from a user’s breath to ensure an accurate reading
from a user’s breath sample.
2. It is recommended that the user does not consume anything for 20 minutes before the breath
test. Following this procedure will eliminate possible misreadings from mouth alcohol and
food.
3. The device can be used with either a sampling cup or a straw adapter and regular drinking
straws (not provided). Set the alcohol test type to either Quick Test Mode or Standard Test
Mode under Advanced Settings (Blow) and install either the sampling cup or straw adapter
onto the device before using.
a. The sampling cup is recommended for Quick Test Mode. The sampling cup should be
cleaned with a non-alcohol disinfectant and wiped dry between each test to prevent
cross contamination or alcohol residue from affecting the next reading.
b. The straw adaptor is recommended for Standard Test Mode. Users should be
provided with individually wrapped regular drinking straws (not provided) to take a
test. Each user should use a new straw for each test and dispose of the used straw
hygienically after testing.
4. After switching on the device, allow 10 minutes for the device to warm up before performing
the Admin Temperature Calibration.