Site preparation checklist
Complete all items in the following checklist before y
ou install the instrument.
IMPORTANT! For installation by a Thermo Fisher
Scientic service representative,
complete, date, and initial the following checklist before the scheduled installation
date. If the site preparation checklist is not complete when the service representative
arrives, the scheduled installation may be postponed.
3 Date Initials Site preparation requirement See page
Customer responsibilities have been reviewed.
4
Personnel have been assigned tasks and responsibilities.
The installation site is
identified and meets the following requirements:
Space and clearance 5
Environmental 7
Electrical 8
Instrument-to-computer connections 10
Safety 12
The instrument was received and inspected as follows:
13
All shipping list items are the items ordered at the time of purchase.
Any damage to shipping containers was reported to the shipping
c
ompany that delivered the instrument.
Any damage or mishandling was recorded on the shipping documents.
All reagents and plates are unpacked and stored as specified.
The installation site is cleared and ready for the instrument installation.
13
The packaged instrument and other shipping containers are moved to the
ins
tallation site.
All materials for installation and operation are available.
14
Register at
thermofisher.com/cloud to access free analysis software and
10 GB of Thermo Fisher Cloud storage.
16
The
QuantStudio
™
3 and 5 Real-Time PCR Systems IT Checklist
(Pub. No. MAN0013738) has been completed and returned.
—
Site preparation checklist
QuantStudio
™
3 and 5 Real-Time PCR Systems Site Preparation Guide | MAN0010405 C.0
3