Instrument-to-computer connections
W
e support the following direct, networked (LAN– local area network), or Cloud
congurations.
Congurations other than those listed are not recommended. Select a
conguration that meets the needs of your laboratory's instrument, software, and
workow requirements.
IMPORTANT! Do not connect both an Ethernet cable and the W
i-Fi module
(Cat. No. A26774) to the instrument. Conguring the instrument for both wired and
wireless connection can interfere with instrument operation.
Note: For detailed information about netw
orking your instrument, see the
QuantStudio
3 and 5 Real-Time PCR Systems IT Checklist (Pub. No. MAN0013738).
Direct configuration option
A computer provided by Thermo Fisher
Scientific with the QuantStudio
Design and Analysis desktop Software
Computertoinstrument connection:
Direct, wired connection between the computer and the instrument
using an Ethernet cable
IMPORTANT! A dir
ect instrumenttocomputer connection
cannot
be combined
with the LAN or the Cloud configurations.
Networked
configuration option
A computer provided by Thermo Fisher Scientific with the QuantStudio
Design and Analysis desktop Software
The computer and instrument must be on the same subnet mask.
ComputertoLAN connection
(select an option)
:
Wired connection to the network using an Ethernet cable –
or
Wireless connection to the network
LAN
LAN
Instrumentt
oLAN connection
(select an option)
:
Wired connection to the network using an Ethernet cable –
or
Wireless connection to the network using the instrument WiFi module
Thermo Fisher Cloud configuration option
Internet access and a computer with the Chrome
w
eb browser to access the Cloud
ComputertoCloud connection
(select an option)
:
Wired connection to the internet using an Ethernet cable –
or
Wireless connection to the internet
LAN
Instrumentt
oCloud connection
(select an option)
:
Wired connection to the network using an Ethernet cable –
or
Wireless connection to the network using the instrument WiFi module
Supported options
for instrument
and computer
connections
Instrument-to-computer connections
10
QuantStudio
3 and 5 Real-Time PCR Systems Site Preparation Guide | MAN0010405 C.0

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems QuantStudio 3 and is the answer not in the manual?

Applied Biosystems QuantStudio 3 Specifications

General IconGeneral
BrandApplied Biosystems
ModelQuantStudio 3
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals