We recommend the use of electrical protective devices to protect the system in
environments with large v
oltage and power uctuations.
Recommended devices
Power line regulator
1.5-kVA power line regulator
Use in areas where the supplied power fluctuates in excess of ±10% of the normal voltage.
Power fluctuations can adversely affect the function of the instrument and computer.
Note: A power line regulator monitors the input current and adjusts the power supplied to the instrument or
computer. It does not protect against a power surge or failure.
Surge protector
10-kVA surge protector (line conditioner)
Use in areas with frequent electrical storms or near devices that are electrically noisy, such as refrigerators, air
conditioners, or centrifuges.
Short-duration, high-voltage power fluctuations can abruptly terminate the function of, and thereby damage the
components of, the computer and the instrument.
Note: A dedicated line and ground between the instrument, computer, and the building’s main electrical service can
also prevent problems caused by power fluctuations.
Uninterruptible power supply (UPS)
1.5-kVA uninterruptible power supply (UPS)
Use in areas prone to power failure.
Power failures and other events that abruptly terminate the function of the instrument and computer can
corrupt data and possibly damage the system.
WARNING! PHYSIC
AL INJURY HAZARD. Do not attempt to lift the UPS unit without assistance of at least
two people. Improper lifting can cause painful and permanent back injury. Refer to the UPS manufacturer
user guide for more information.
IMPORTANT! A UPS pr
ovides power for a limited time. It is meant to delay the effects of a power outage, not to
serve as a replacement power source. In the event of a power loss, power off the instrument and computer unless
you expect to regain power within the battery life of the UPS.
Electrical
protective devices
Site requirements
QuantStudio
3 and 5 Real-Time PCR Systems Site Preparation Guide | MAN0010405 C.0
9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems QuantStudio 3 and is the answer not in the manual?

Applied Biosystems QuantStudio 3 Specifications

General IconGeneral
BrandApplied Biosystems
ModelQuantStudio 3
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals