BASE KEYWORD DESCRIPTION VALUE
OUTPut
:[STATe] <value>
:[STATe]?
Set Output State ON, OFF
MODE <value>
:MODE?
Set Output Mode MANual, LIST, PULSe, STEP
VOLTage:AC <value>
VOLTage:AC?
Set AC Voltage 0.0 – 310.0 VAC
:VOLTage:DC <value>
:VOLTage:DC?
Set DC Voltage 0.0 – 420.0 VDC
:FREQuency <value>
:FREQuency?
Set Frequency 5.0 – 1200 Hz
:CURRent[:LIMit]:HIGH
<value>
:CURRent[:LIMit]:HIGH?
Set Current High Limit 0.05 – 12.50 A*, 0 = OFF
:PROTection:STATe? Query Protection State
FAN_FAIL, SOFT_S, OPP, OCP,
OVP, OPP_PEAK, OCP_PEAK,
OVP_PEAK, INT_TEST_FAIL,
CALIBRATION_ERROR,
Limit_Fail, DD_OCP, DD_UVP,
MCU_FAIL, DA_OTP, PFC_OVP,
AC_LVP, AC_OVP, SET_FAIL,
OUTPUT_SHORT, RCP, RCP_
PEAK, DD_FAIL
:PROTection:CLEar Clear Protection State
7.4 SCPI Command TREE
Program Commands and Companion Queries
These commands are used to modify individual test parameters within each le and/or sequence. Many
of these commands require a parameter value to be included with the command. The companion query
command will read the parameter. The writing of the parameter requires that the unit not be included
with the value and only the numeric value should be included with the command. Also, when the query
commands are sent, the response will not include the unit characters.
7.4.1 - Output Subsystem