MEASure:ALL?
Reads the active data being displayed on the LCD display while the test is in process. This command
will also read the last data taken when the test sequence has completed. Each parameter is
separated by commas and includes frequency value, voltage value, current value, power value,
peak current value, power factor value and timer metering. The syntax for the command response is
<V>,<VAC>,<VDC>,<A>,<AAC>,<ADC>,<F>,
<P>,<PF>,<AP>,<Q>,<CF>,<VA>. Each meter will contain only the value and not the units. Current and
peak current are displayed in amps while power is displayed in Watts.
RESult:ALL?
Reads the results for an individual sequence. The sequence number has to be selected rst by sending
RESult:Sequence <numerica_value>, this will select the appropriate sequence number. Each results
parameter is separated by commas and includes all metering data. The syntax for this command response
is <Voltage>,<Vac voltage>,<Vdc voltage>,<Current>,<AC
MEASure:FREQuency?
Reads the active frequency value being displayed while a test is in process.
MEASure:VOLTage:AC?
MEASure:VOLTage:DC?
Reads the active voltage value being displayed while a test is in process.
MEASure:CURRent?
MEASure:CURRent:AC?
MEASure:CURRent:DC?
Reads the active current value being displayed while a test is in process.
MEASure:APEAK?
Reads the active peak current value being displayed while a test is in process.
MEASure:POWer?
Reads the active power value being displayed while a test is in process.
MEASure:PFACtor?
Reads the active power factor value being displayed while a test is in process.
MEASure:TIMe:DWELl?
Read the active timer meter value being displayed while a test is in process.
METer POWer
Selects the metered value that is displayed while a test is in process. To change meters send a new
command with name of the meter from the list below:
FREQuency|CURRent|POWer|PFACtor|APEAK|REACtive|CREStfactor|APParent|VAC|VDC|AAC|ADC