AquaLab 7 VERIFICATION AND CALIBRATION
second verification standard should be within ±0.003 a
w
.
7. If either of the verification standards is not correct, it is prob-
ably due to contamination of the sensor chamber. For cleaning
instructions, see Chapter 6. After cleaning, repeat verification
from step two.
8. If you are consistently getting readings outside the water ac-
tivity of your first verification standard by more than ±0.003
a
w
, a linear offset has probably occurred. In this case, adjust
the reading to match the correct verification standard value as
outlined in the next section.
Figure 3: Verification Standard Flowchart
Note: See the verification section in the operator’s manual for de-
tails. The Measure Verification Standard flowchart is a graphical
representation of the Verification of Calibration directions.
37

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the AquaLab 4TE Series and is the answer not in the manual?

AquaLab 4TE Series Specifications

General IconGeneral
Accuracy±0.003 aw
Temperature Control15 to 50 °C
Measurement Time5 minutes
Measurement PrincipleChilled Mirror Dewpoint
Resolution0.0001 aw
Temperature Accuracy±0.1 °C
CommunicationUSB, RS-232
DisplayLCD
Measurement Range0.03 to 1.00 aw
Operating Temperature4-50 °C (39-122 °F)
Data Storage8000 readings
Power Supply50/60 Hz
Storage Temperature-20 to 50 °C

Related product manuals