AquaLab 9 TAKING A READING
and 50
C. Between these ambient temperatures, AquaLab will mea-
sure samples of similar temperature quickly and accurately. The
AquaLab Series 4TE and 4TEV have temperature control capabili-
ties that enable them to read samples at temperatures different from
ambient temperature, but no higher than 50
C.
If a sample has a water activity lower than about 0.03, AquaLab will
display the < symbol in the Measurement window to notify you that
your sample is too dry to be accurately measured by the AquaLab.
If you know that your sample water activity is above what the screen
is telling you, your instrument sensors may have been contaminated
and will need to be cleaned (see Chapter 6) or serviced (see Chapter
13).
49

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the AquaLab 4TE Series and is the answer not in the manual?

AquaLab 4TE Series Specifications

General IconGeneral
Accuracy±0.003 aw
Temperature Control15 to 50 °C
Measurement Time5 minutes
Measurement PrincipleChilled Mirror Dewpoint
Resolution0.0001 aw
Temperature Accuracy±0.1 °C
CommunicationUSB, RS-232
DisplayLCD
Measurement Range0.03 to 1.00 aw
Operating Temperature4-50 °C (39-122 °F)
Data Storage8000 readings
Power Supply50/60 Hz
Storage Temperature-20 to 50 °C