AquaLab 4TE Series - Water activity readings on verification standards are too high;low and a linear offset adjustment cannot be made any higher;lower; Message on screen displays a warning that the sample is too hot; DUO Model-Test was run with wrong model; DUO Model-%Moisture Content displayed is not correct

AquaLab 4TE Series
116 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AquaLab 12 TROUBLESHOOTING
13. PROBLEM:
Ran test with wrong model.
SOLUTION:
1. On the measurement screen, toggle to the correct model using
the up and down arrow keys. The moisture content value will
be updated to correspond with the model selected.
2. If the correct model is not available, the model may not be
loaded on the instrument.
(a) To determine which models are loaded on the instrument,
cycle to the menu tab, select Moisture Content and then
the loaded models will appear.
3. If the correct model is not available, load the appropriate model
using AquaLink 4 Software. The AquaLab DUO can hold a
total of 100 models at any one time. You may need to re-
move a model using the Software or use the delete option in
the %Moisture Calibration menu before you can add a new one.
Any model that is removed from the instrument with AquaLink
4 will be stored and may be reloaded again later.
14. DUO PROBLEM:
Moisture Content displayed is not correct.
SOLUTION:
1. Model selected may not be correct for the product being tested.
(a) Toggle through the available models to find a more appro-
priate model.
(b) If the model is correct but not giving correct moisture con-
tent values it may be necessary to generate a new model
for the product or update an existing model. For informa-
tion about generating a model, contact Decagon Devices
for updating a model, refer to Chapter 10: DUO Opera-
tion.