2-10
Table 2.2 - Power Factor/Phase Operating Procedure
______________________________________________________________________________
STEP OPERATOR ACTION RESULTS
______________________________________________________________________________
1. Apply power to PMC After successful completion of Function Tests the PMC will
momentarily display PASSED FUNCTION TEST. The PMC will
return to RESET mode and display DC, 60 Hz, or 400 Hz? The
RESET, STBY, and NORMAL LED's (and the STORE LED if
user settings are presently stored in the user memory) will be
illuminated.
2. Press 60 Hz RESET LED will turn off, the 60 HZ and FREQUENCY LED's
will turn on. The PMC now displays FREQUENCY 60.00 Hz.
(The Frequency may now be modified or a new value entered).
3. Press Power Factor FREQUENCY LED will turn off, the POWER FACTOR LED
will turn on and the PMC now displays POWER FACTOR 1.000
(Power Factor may now be modified or a new value entered).
4. Press Voltage VOLTAGE LED is illuminated and the PMC now displays
VOLTAGE 15.00V (15 Volts rms is the lowest range available
for Power Factor/Phase output mode).
5. Press 1 The PMC displays ENTRY = 1
6. Press 2 The PMC displays ENTRY = 12
7. Press 0 The PMC displays ENTRY = 120. (The MINUS/BACKSPACE
key may be used to erase incorrect entries).
8. Press V/A Press the unit key (V/A) to complete the entry of the voltage
setting. The PMC now displays VOLTAGE 120.0 V (The
voltage may now be modified using the rotary Knob or the
Up/Down modify keys.).