X-ray Lock
Alert
X-RAYS
LOCKED
Tones will be
the same as
the actual
operation
tones
When the trigger is activated while the device is
in the X-rays locked mode and operation is
attempted, various alert messages will be
displayed. For example:
X-RAYS LOCKED
,
SIMULATION ONLY
,
SIMULATION COMPLETE
.
Toggle between
LOCKED
and
UNLOCKED
by
pressing and holding down hold down the
Select
( ) button and then press the
Patient
( ) button.
Duty Cycle
Exceeded
Alarm
COOLING
DOWN
[count down]
Double tone
at the start
and end of
the cool down
cycle
If the operator presses the trigger to activate the
READY
state before the duty cycle time, the
device will display the warning message, and a
countdown timer will be displayed showing the
duty cycle time remaining. This will lock the
device and the message will be displayed until
the duty cycle is complete. The device will then
return to the normal operation mode.
Low Battery
Alarm
RECHAR
GE
HANDSE
T
5 audible
tones
Ensure the Handset is locked in place.
The
RECHARGE HANDSET
alert terminates after
five seconds and goes into auto-shutdown.
Replace the current Handset with a freshly
charged Handset.
NOTE: If battery voltage is lower than required
for the X-ray exposure, the device will not allow
the exposure.