A warning
tone
Activates if trigger is released before the timed
X-ray exposure finishes.
This condition is cleared by pressing any button
on the user interface panel or by tuning the
power off then on.
Depress the trigger for the duration of the
exposure time.
System
Failure
Alarm
SYSTEM
FAILURE
CALL
MANUFACTU
RE
Descending
tone
Consult Section 6.2, Troubleshooting (6.2.8) or
an Aribex authorized service center.
Handset
Failure
Alarm
HANDSE
T
FAILURE
5 audible
tones
Consult Section 6.2, Troubleshooting (6.2.12) or
an Aribex authorized service center.
NOMAD Pro should not be operated if it has been dropped or performance degrades; it
should be returned to Aribex for a safety check.