CARE AND PREVENTIVE MAINTENANCE46
Caregiver obligations - before every
use or every week (if in long-term
use)
Visually check the Auralis pump for damage
&KHFNWKDWWKHFRQWUROSDQHOLV¿UPO\DI¿[HGLV
undamaged and legible.
Check all electrical connections and power
cable
for signs of excessive wear or damage.
Check the connector socket for signs of excessive
wear or damage.
Visually check the Auralis mattresses and
Auralis seat cushion for damage
Check that the tubing is not damaged and that
there are no obvious cracks or breaks to the
connector.
Remove the top cover and
check for signs of wear
or any tears.
Visually check the Auralis mattress
extensions for damage
Remove the top cover and check for signs of wear
or any tears.
Inspect the foam core for any signs of damage,
soiling or staining.
If any part is damaged or missing DO NOT use the
product.
Perform a full functionality test on the Auralis
system (6 steps)
1. Connect the tube-set to the Auralis pump. Make
sure the tube-set clicks into place.
2. Connect the power cable to a power source.
The Auralis pump sounds with a start-up tone
and a self-diagnostic check runs for 10 seconds.
Once the check is completed the Auralis pump beeps
twice and is ready for use.
3. Press the Run/standby button to start the Auralis
pump.
4. To check the alarm functionality, disconnect
the tube-set from the pump while the system is
switched on. After 30 seconds, two green alarm
panel indicators on the front of the pump turn
yellow, the Low pressure indicator turns yellow
and an audible alarm sounds.
5. Reconnect the tube-set to the pump to deactivate
the alarm.
6. If the functionality test is not completed, call
TXDOL¿HGVHUYLFHSHUVRQQHO
Caregiver obligations - after each
patient
Clean and disinfect
• The Auralis system has to be cleaned and
disinfected after each pati
ent, see section
³&OHDQLQJDQGGLVLQIHFWLRQLQVWUXFWLRQV´R
n page
40.
Visually check the Auralis mattresses and
Auralis seat cushion for damage
Check that all seams and the zips are undamaged
and not loose. Check that the zip puller is not
missing.
&KHFNWKDWWKH¿[LQJDQGVHFXULQJVWUDSVDUH
secure and not damaged
Check that all cell fasteners, loop straps and
Heelguard™ cell straps, are correctly connected
to the
Auralis mattress base and are not loose or
damaged.
Check that all cell hooks are attached.
See Figure 63
Figure 63
If any part is damaged or missing DO NOT use the
product.