Electrical
Maximum sound power level
Environment
Duty cycle Max volts Max amps
Mast Lift Actuator (sealed electro-hydraulic unit) 10% (2 min/18 min) 24 V 20 A
‘V’ Chassis Actuator (electro-mechanical unit) 10% (2 min/18 min) 24 V 8 A
In accordance with ISO 3746
(dB re 1pW ± 3dB)
74 dB
Operating, transport and storage
Temperature +10 °C to + 40 °C (+50 °F to +104 °F) Operation
-20 °C to +70 °C (-4 °F to +158 °F) Transport
-20 °C to +70 °C (-4 °F to 158 °F) Storage
Relative humidity range 30% to 75% Operation
10% to 80% including condensation Transport and Storage
Atmospheric pressure 800 hPa to 1060 hPa Operating
500 hPa to 1100 hPa Transport
500 hPa to 1100 hPa Storage
End of Life Disposal
All batteries in the product must be recycled separately. Batteries are to be disposed in accordance with
national
or local regulations.
Lift systems having electrical and electronic components or an electrical cord should be disassembled and
recycled per Waste of Electrical and Electronic Equipment (WEEE) or in accordance with local or national
regulation.
Components that are primarily be made up of different kinds of metal (containing more than 90% metal by
weight) for example sling bars, rails, upright supports, etc., should be recycled as metals.
Technical specification