15A3356O 11
If NO is pressed, the following prompt will be displayed:
Main Menu
Pipette Calibration?
If YES is pressed to continue with the calibration, there is no
certainty that the results the PCS
obtains will meet the system
specifications.
The display will prompt to enter the Lot Code:
Enter Reagent Lot
Code: _
4.3.2 Enter the Lot Code. The Lot Code is shown on the Reagent Vials,
Reagent Kit insert, and on the exterior of the Reagent Kit box. The
Instrument maintains a library of up to 6 previously entered Lot
Codes. If this Lot Code matches one previously entered, proceed to
Section 4.3.3.
If the Lot Code does not match, the display will prompt:
Rgt Lot Code XXXXX
New Lot? (Y/N)
If the Lot Code was entered incorrectly, press NO and enter the
correct Lot Code.
Press YES if entering a new Lot Code. The Group Numbers found
on the correct side of the Lot Code Card for the instrument type will
need to be entered. Press ENTER after entry of each group of
digits. At the next prompt, enter the month and year of the
expiration date of the Reagent Kit as indicated on the Lot Code
Card in the format mmyy (e.g. 0811 for August 2011).
The following prompt will be displayed:
Lot Code: XXXXX
Correct? (Y/N)
Press YES to confirm.