15A3356O 35
Access this option using the Setup Menu. (Refer to Section 6.1,
“Instrument Setup Menu.”) See Appendix C for formulas.
Press YES at the prompt:
Setup Menu
Choose Nomenclature?
The following prompt will be displayed:
Nomenclature: XXX
Change? (Yes/No)
Press NO to keep the nomenclature as displayed. Press YES to change
the nomenclature to the alternative selection. See Appendix B for example
of Printout.
6.9 Display Uncertainty (available in version 7A2.2XX or newer firmware)
The Combined Expanded Uncertainty of the mean for each calibration can
be calculated by the PCS and displayed on the print out if desired. The
calculation is made using data from the calibration including mean volume,
%CV, number of data points and standard uncertainties, u
i
for each PCS
Range Solution. See Appendix C for calculation method. Access this
option using the Setup Menu. (Refer to Section 6.1, “Instrument Setup
Menu.”)
Press YES at the prompt:
Setup Menu
Display Uncertainty?
To skip this setup option, press NO.
If YES is pressed, the following prompt will be displayed:
Uncertainty: Omit
Change? (Yes/No)
or
Uncertainty: Include
Change? (Yes/No)
These prompts indicate whether the uncertainty is currently omitted from,
or included in, the printout.
Press NO to keep the Uncertainty setting as displayed. Press YES to
change the Uncertainty setting to the alternative selection. See Appendix
B for an example Printout.