15A3356O 18
Press NO. If a given data point from the statistical summary needs to be
deleted, press YES. At the prompt:
Enter Pt # to Delete
_
Enter the Sample # of the data point to be deleted. At the prompt:
Delete More Points?
(Yes/No)
Press YES to continue to delete data points. Any number of data points
may be deleted. The deleted data points will be printed on the Printout, but
an asterisk will appear beside them, indicating that they are not included in
the statistical summary. When all desired data points are deleted, press
NO at the prompt:
Delete More Points?
(Yes/No)
The corrected statistics will be printed.
4.10 Performing Additional Calibrations In One Vial
The number of data points that can be collected with each Blank Vial will
depend on the volume of Sample Solution added and range from 1 to a
maximum of 40. (Refer to Appendix F.)
Each time a pipette calibration run is completed, the Instrument will
display the remaining number of samples of the selected volume that can
be dispensed into the vial. Additional calibration runs may be completed
within the vial’s allowed capacity. The options include performing
calibrations:
(a) at the same volume
(b) at different volumes but within the same range of Sample
Solution
(c) at different volumes and with a different range of Sample
Solution
IMPORTANT When calibrating pipettes from more than one range in a given vial, always use
the range of Sample Solution which is appropriate for the volume of pipette being
calibrated.
Following the completion and printing of the first calibration run, the
following prompt will be displayed: