Enter the Operator ID (not exceeding 20 digits or characters) for
this calibration and press ENTER. This ID appears on the Printout.
For alphabetic character entry, see Appendix E. If the ID is entered
incorrectly, press CLEAR and enter the correct ID.
4.5 Entering Pipette ID Number
At the prompt:
Enter Pipette ID:
_
Enter the Pipette ID (not exceeding 20 digits or characters) for this
calibration and press ENTER. This ID will be printed on the PCS Printout
for pipette cross-referencing, and as a hard copy record. For alphabetic
character entry, see Appendix E.
The Instrument will print the information just entered and display:
*** Printing ***
Please Wait
Followed by:
Enter Pipette
Volume: _ uL
4.6 Entering Pipette Volume
Enter the volume (in microliters) to be dispensed, and press ENTER. The
CLEAR key can be used to correct an entry error prior to pressing the
ENTER key.
NOTE: This volume must be entered in microliters ranging from 0.1 μL through 5000 μL.
The number of samples that can be added at this pipette volume in
this vial will be displayed:
XX Left this Vial
Proceed? (Yes/No)
Press YES. The following prompt will be displayed:
Range X
Please Wait
The Instrument will take a final Blank reading, and then prompt: