15A3356O 19
Another Cal w/ Vial?
(Yes/No)
Press YES to continue. The next prompt will instruct to enter an
Operator ID:
Enter Operator ID:
1
Enter the Operator ID (not exceeding 20 digits or characters) for this calibration and press
ENTER. This ID appears on the Printout. For alphabetic character entry, see Appendix E.
If the ID is entered incorrectly, press CLEAR and enter the correct ID.
The next prompt will instruct to enter the Pipette ID:
Enter Pipette ID:
_
Continue the calibration of the second pipette from Section 4.5,
“Entering Pipette ID Number.”
When the vial's maximum capacity is reached, the following prompt
will be displayed:
Result #XX XXX.X
along with the alternating prompts:
End of Vial
and
Use New Vial? (Y/N)