15A3356O 22
Table 1: Artel’s suggested initial tolerance limits at common calibration volumes.
Pipette Volume, µL Relative Error Absolute Error
Nominal Setting
Inaccuracy ±% %CV Inaccuracy ± µL STD µL
2
2.0
1.0
0.5
0.2
2.0
4.0
8.0
20.0
1.0
2.0
4.0
10.0
0.04
0.02
2.5
2.5
2.0
1.0
0.5
0.25
2.0
2.5
5.0
10.0
20.0
1.0
1.3
2.5
5.0
10.0
0.05
0.025
10
10
5
2
1
2.0
4.0
10.0
20.0
1.0
2.0
5.0
10.0
0.20
0.10
20
20
10
5
2
2.0
4.0
8.0
20.0
1.0
2.0
4.0
10.0
0.4
0.2
50
50
25
10
5
2.0
4.0
10.0
20.0
1.0
2.0
5.0
10.0
1.0
0.5
100
100
50
25
10
2.0
4.0
8.0
20.0
1.0
2.0
4.0
10.0
2.0
1.0
200
200
100
50
20
2.0
4.0
8.0
20.0
1.0
2.0
4.0
10.0
4.0
2.0
500
500
250
50
2.0
4.0
20.0
1.0
2.0
10.0
10.0
5.0
1000
1000
500
250
100
2.0
4.0
8.0
20.0
1.0
2.0
4.0
10.0
20.0
10.0
2000
2000
1000
500
200
2.0
4.0
8.0
20.0
1.0
2.0
4.0
10.0
40.0
20.0
2500
2500
1000
500
2.0
5.0
10.0
1.0
2.5
5.0
50.0
25.0
5000
5000
2500
1000
500
2.0
4.0
10.0
20.0
1.0
2.0
5.0
10.0
100.0
50.0