15A3356O 24
Main Menu
Pipette Calibration?
Press NO or MENU SCROLL.
Main Menu
Instr. Calibration?
Press YES. The following prompt will be displayed:
Instrument Cal. Menu
Perform Instr. Cal.?
Press YES to continue with an instrument calibration.
The instrument will then check the temperature, and if it is outside
the specified range for proper operation (18 °C to 28 °C), the
following prompt will be displayed:
Temp. out of range
Continue? (Y/N)
If NO is pressed, the following prompt will be displayed:
Main Menu
Pipette Calibration?
If YES is pressed to continue with the calibration, there is no
certainty that the results the PCS
obtains will meet the system
specifications.
5.2.2 At the following prompt:
Enter Operator ID:
1
Enter the Operator ID (not exceeding 20 digits or characters) for
this calibration and press ENTER. This ID appears on the Printout.
For alphabetic character entry, see Appendix E. If the ID is entered
incorrectly, press CLEAR and enter the correct ID.