Preventative Maintenance Inspection Form
DFU-0276-4r0_fmt_en-US PM – 3/8
Setup:
To aid in testing the Synergy
RF
system, obtain an Apollo
RF®
Service probe, an Electrical Safety Analyzer,
and an Electrosurgical Analyzer.
A Calibrated Electrical Safety Analyzer will be required to complete the electrical safety test.
Procedure:
Indicate a PASS/Fail for each item. Any failure indicates that the device will need to be sent to Arthrex
for evaluation.
Visual Inspection
Initials
Action
Pass/Fail
Visual Inspection: Visually inspect the Console and foot pedal to ensure that the
cables, connectors, contacts, and display are not damaged or worn. If fluid or
particles have splashed on the display, clean with a microfiber cloth by gently
wiping in a circular motion.
Preparation
Initials Action
Present/Not
Present
Make sure the Synergy
RF
Console screen appears when the AC main power
switch is activated
Connect Service Probe
Warnings
Initials Action
Present/Not
Present
Check that error messages are not present. [See Appendix]
Any error message need to be evaluated and corrected before moving forward with this Procedure.
Please see the Appendix at the end of this document for any error message that is displayed.

Table of Contents

Other manuals for Arthrex SynergyRF AR-9800

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arthrex SynergyRF AR-9800 and is the answer not in the manual?

Arthrex SynergyRF AR-9800 Specifications

General IconGeneral
BrandArthrex
ModelSynergyRF AR-9800
CategoryMedical Equipment
LanguageEnglish

Related product manuals