Preventative Maintenance Inspection Form
DFU-0276-4r0_fmt_en-US PM – 4/8
Touch Screen
Initials Action Pass/Fail
Make sure the ablate number increases when pressing the yellow ablate "Up
Button" [3]. The ablate power [1] will increase in increments of 1.
Make sure the ablate number decreases when pressing the yellow ablate
"Down Button" [3]. The ablate power [1] will decrease in increments of 1.
Make sure the coag (coagulation) number increases when pressing the blue
coag "Up Button"[4]. The coag power [2] will increase in increments of 1.
Make sure the coag number decreases when pressing the blue coag "Down
Button"[4]. The coag power [2] will decrease in increments of 1.
Make sure you are able to press the "Menu" button.[5]
Once on the menu, make sure you are able to navigate through the menu
settings without any problems.