EasyManua.ls Logo

ASTECH LDM41A - 10 List of Commands

ASTECH LDM41A
55 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
List of Commands LDM41A and LDM42A Manual
Page 48 ASTECH GmbH
10 List of Commands
Table 9 : List of commands
Command
Description
DT
Starts distance tracking
DS
Starts distance tracking (< 7 m)
DW
Starts distance tracking on white target at 10 Hz
DX
Starts distance tracking on white target at 50 Hz (only
LDM42A)
DF
24
Starts remote-triggered single distance measurement
(single shot)
DM
Starts single distance measurement (single shot)
TP
Queries inner temperature
SA
Queries / sets floating average value (1...20)
SD
Queries / sets output format (dec/hex/sig)
ST
Queries / sets time to measure (0...25)
SF
Queries / sets scale factor
SE
Queries / sets error mode (0, 1, 2)
AC
Queries / sets alarm center
AH
Queries / sets alarm hysteresis
AW
Queries / sets alarm with
HO
25
Queries / sets temperature for heating on
HF
25
Queries / sets temperature for heating off
RB
Queries / sets beginning of range (4 mA)
RE
Queries / sets end of range (20 mA)
RM
Queries / sets removal measurement parameters
TD
Queries / sets trigger delay and level
TM
Queries / sets trigger mode and level
BR
Queries / sets baud rate
AS
Queries / sets auto start
24
Trigger function is not available for devices with internal heating (LDM41A-h and LDM42A-h)
25
Only for devices with internal heating (LDM41A-h and LDM42A-h)