EasyManua.ls Logo

Avady Propilot Salt - Calibrating the ORP Probe; Activation;Deactivation of Ph Regulation; Manual Injection

Avady Propilot Salt
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13
4.6.18. Calibrating the ORP probe
1) Open the 475 mV ORP standard solution.
2) Turn off the filtration (and therefore the electronics unit).
3) If the probe is already installed:
a) Remove the probe from the probe holder, without disconnecting it.
b) Insert the cap provided into the probe holder.
If the probe is not already installed:
Connect the probe to the electronics unit.
4) Turn on the electronic unit.
5)
6) Follow the instructions below:
Insert the probe into the ORP calibration solution, then wait 5 minutes.
Do not touch the probe.
(Wait a few seconds)
a) Rinse the probe under running water, then leave it to drip
-
dry
without wiping it
.
b) Install the probe into the probe holder.
or
Perform the navigation again with the above instructions, several times if
necessary.
If calibration still fails, replace the probe then perform another calibration.
4.6.19. Activation/deactivation of pH regulation
Menu Possible settings Default setting
ON
OFF
ON
4.6.20. Manual injection
Menu Functions Possible settings Default setting Instructions
Priming of the
peristaltic pump
and filling of semi-
rigid tubing.
pH corrector
injection.
Means of checking
the correct
operation of the
peristaltic pump.
30
seconds to 10
minutes in 30-
second
increments.
1 min.
To start injecting:
Confirm the duration
setting.
(The peristaltic
pump is running, and the
timer countdown is
displayed in real time.)
To pause, and to restart
the injection:
Press OK.
To stop injection:
Press .
OK
OK

Table of Contents