15. Maintenance and repairs of small defects
1. A moisture analyser should be kept clean.
2. Take care that no dirt gets between the casing and the pan. If found any, remove
the pan (lift it up), remove dirt and then replace the pan.
3. In case of improper operation caused by a short-lasting power supply decay, unplug
the moisture analyzer from the mains and then plug it again after few seconds.
4. It is forbidden to make any repairs by unauthorised persons.
5. To repair the moisture analyzer, please contact an authorised service centre.
Moisture analyser can be sent for repair as messenger delivery only in original
package, if not, there is a risk of damaging the moisture analyzer and loosing
guarantee.
Measuring problems:
Problem
Solution
A sample burns down
Reduce temperature
Use glass fibre filter on the top of the sample
Reduce sample quantity and distribute it uniformly
Drying lasts too long
Increase temperature
Reduce sample mass
A sample loses weight
before measurement
Take out the pan and put a sample outside the chamber
A sample is liquid or paste
Use glass fibre filter
A sample does not consist of
enough volatile matters
Enlarge a sample